Fiber Optic Adaptive Probe as a Cuvette Substitution for UV-Vis Spectrophotometer (Genesis 10S)

Fredy Kurniawan, Edi Suharmanto

Abstract


Fiber optic adaptive probe UV-Vis for Genesis 10S has been fabricated to substitute cuvette for real time analysis. Most of standard spectrophotometers UV-Vis (the common type) use cuvette to place sample which is being analyzed. The light pass through the sample in the cuvette to detector then the intensity of the absorption of the solution is measured. Most of the cuvette system requires to transferring sample solution to cuvette from its original vessel, so the real time analysis cannot be conducted. The probe with special design using fiber optic cable FT-Au tonics 420-10 has been made to substitute the cuvette system. The light from the source moved inside fiber optic cable, which has been cut in the middle of the cable (the sample can passes through at this gap), to the detector. The end (entrance and exit) of the fiber optic was designed so that it fit to the compartment of the Genesis 10S spectrophotometer without any modification from the standard instrument condition. The spectrum results from the probe in varies gap, in the range 0.2 – 1 cm, was observed. The gap in the middle of the fiber optic was equal to the length of the pathway in the cuvette system. Rhoda mine B was used as the sample with variation concentration 4.175 × 10-5 M, 3.34 × 10-5 M, 2.5 × 10-5 M, and 1.67 × 10-5 M. The spectrum result of adaptive fiber optic probe was compared to the spectrum of standard cuvette system. The optimum result was provided using 0.2 cm gap. The maximum wavelength absorbance of adaptive fiber optic probe was at 533 nm which similar to the cuvette system

Keywords


spectrophotometer probe; real time analysis; fiber optic; Rhoda mine B; in-situ

Full Text:

PDF

References


M.H. Mulja, and Suharman, Analisis Instrumental, Airlangga University Press, Surabaya, 1995.

A. Didit, Pembuatan Probe Sensor Serat Optik Untuk Mengukur Derajat Keasaman (pH) Menggunakan Violet Sebagai Dye Indikator. Jurusan Fisika Institut Pertanian Bogor, Bogor, 2006.

T.Y. Hidayat, Penggunaan Serat Optik Pada Spektrofotometer UV-Vis untuk Penentuan Cuplikan Secara In-Situ.Jurusan Fisika-FMIPA, Institut Teknologi Sepuluh Nopember, Kampus ITS Sukolilo, Surabaya, 2013.

C.W. Brown, dan J. Lin, Interfacing a serat-optik probe to adiode array UV-visible spectrophotometer for drug dissolution tests, Appl. Spectroscopy. Wenheim, Jerman, 1993, 615-618.

Y. H. Pramono, dan AY. Rohedi, Aplikasi Directional Coupler Serat Optik sebagai Sensor Pergeseran, Jurusan Fisika-FMIPA, Institut Teknologi Sepuluh Nopember, Kampus ITS Sukolilo, Surabaya, 2008.

M.J. Snare, F. E. Treloar, K. P. Ghiggino, and P. J. Thistlethwaite, The Photophysics of Rhodamine B. Physical Chemistry Department, Melbourne University, Australia, 1981.

P. Wilhelm, dan D. Stephan, Photodegradation of rhodamine B in aqueous solution via SiO2TiO2nano-spheres. Technische Universitat Munchen, Garching, Germany, 2007.

S. Bakkialakshmi, P. Selvarani, S. Chenthamarai, Fluorescence quenching of Rhodamine B base by two amines, Department of Physics, Annamalai University, Annamalainagar 608 002, Tamil Nadu, India, 2013.

J. Alarcon, Fiber Optic Devince For Sensing Analysis and Metode Of Making Same, GooglePatent US 7787923 B2, 2010.




DOI: http://dx.doi.org/10.12962/j20882033.v25i2.536

Refbacks

  • There are currently no refbacks.


Creative Commons License

IPTEK Journal of Science and Technology by Lembaga Penelitian dan Pengabdian kepada Masyarakat, ITS is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Based on a work at https://iptek.its.ac.id/index.php/jts.