Blood Pressure Monitor Design Using MPX5050GP Pressure Sensor and Visual C# 2010 Express
Abstract
Blood pressure monitor (BPM) using oscillometric method with MPX5050GP pressure sensor has been successfully fabricated and calibrated. Oscillometric method is measurement method to observe cuff pressure oscillations. In this study, the oscillometric waveform measured by fabricated BPM used Arduino Uno which processes pulse rate, systolic and diastolic pressures. Measurement results of pulse rate, systolic and diastolic pressures will be displayed and recorded in application designed using Visual C# 2010 Express. Fabricated BPM capable of measuring pulse rate and blood pressure in the range of 40-260 mmHg with ratio of systolic and diastolic pressures were achieved 0.48 and 0.78, respectively. The system has an average error of systolic pressure, diastolic pressure, and pulse rate were -4.38 mmHg (2.72%), 0.10 mmHg (5,55%), and 4.48 bpm (6.07%).
Keywords
Full Text:
PDFReferences
Oktavianto, H., Arif, T., Proceedings of Industrial Electronics Seminar, Surabaya, Indonesia, 57 (2002) p.1–4.
Newell, S.R., Master's Thesis, Department of Health Science and Technology, Aalborg University, Denmark, 2013.
Ball-llovera, A., Rey, R.D., Ruso, R., Ramos, J., Batista, O., Niubo, I., Proceedings of 25th Annuals Intemational Conference of the IEEE EMBS, Cancun, Mexico, 2003, p.3173–3175.
Jaafar, R., Desa, H.M., Mahmoodin, Z., Abdullah, M.R., Zaharudin, Z., International Conference On Instrumentation, Communication Information Technology and Biomedical Engineering, Bandung, Indonesia, 2011, p.265 – 269.
Drzewiecki, G., Hood, R., Apple, H., Annals Biomedical Enggineering 22 (1994) 88–96.
Bronzino, J.D., Medical Devices and Systems, 3rd ed., CRC Press, USA, 2003, p.55-6.
Arteta, Domingos, Pimentel, Santos, Chiffot, Springer, Raghu, Clifford, Proceeding of 2nd International ICST Conference on Wireless Mobile Communication and Healthcare - Mobihealth, Kos Island, Greece, 2011, p.335-342. http://www.robots.ox.ac.uk/~gari/papers/BP_Arteta_et_al.pdf, 2011
Moraes, J., Cerulli, M., Ng, P., IEEE Computers in Cardiology. 27 (2000) 211–214.
Porr, B., BPM Biosignals, http://biosignals.berndporr.me.uk/doku.php?id=nonelec:pressure, 2014.
Datasheet MPX5050GP, 2010.. Freescale Semiconductor, http://www.farnell.com/datasheets/673750.pdf, 2010.
Huang, J.-T., Lee, K.-Y., Chiu, M.-C., Intech (2008), pp. 444.
Chua, C.., Hin, S.M., Freescale Semiconductor Application Note, http://bioingenieria.edu.ar/academica/catedras/bioingenieria2/archivos/datasheet/an1571.pdf, 2005